Experimental Analysis of Errors Induced in Surface Temperature Measurement on Aerospace Vehicle

DOI : 10.17577/IJERTCONV1IS03017

Download Full-Text PDF Cite this Publication

Text Only Version

Experimental Analysis of Errors Induced in Surface Temperature Measurement on Aerospace Vehicle

N.Varalakshmi, K.Gobi, M.R.Satyanarayana and A.V. Sudarsana Devi Defence Research & Development Laboratory, Hyderabad 58 Email: varakutti@gmail.com

AbstractTemperature is inherently a transient phenomenon in which heat is constantly flowing from a hotter medium to a colder medium by conduction, radiation and convection. Insulation barriers such as walls, shields, etc., simply slowdown the rate of heat flow, but do not stop it. Accurate measurement of surface temperature is extremely difficult to accomplish because, regardless of type of temperature sensor used for measurement, the temperature sensed is the temperature of the sensor- not the temperature of the surface. This paper presents the possible sources of error in surface temperature measurement using

  1. SURFACE TEMPERATURE MEASUREMENT

    1. Temperature Measurement Chain

      T/C

      S/C

      DAS

      thermocouples and experimental analysis of error due to the cables and connectors used for interfacing thermocouples to signal conditioner.

      T/C Wire Extension Wire

      Signal Cable

      T/C Connector

      Keywords-Aerospace enginest; surface temperature measurement; cement- on thermocouples; sources of error; experimental analysis

      1. INTRODUCTION

        The continuous development and evaluation of Aerospace components, systems and engines require accurate measurement of surface temperature during testing to aid design. Thermocouples are preferred for this purpose due to its wide ranges, availability of different types, ease handling and simple in construction [1]. As thermal conduction error of probe type thermocouple is dominant, it is less preferred for surface temperature measurement. Thin film thermocouples are mostly used in Aerospace applications due to its fast response time and less perturbation of surface on which temperature to be measured[2].Thin film thermocouples provide a more accurate measurement of the actual surface temperature due to its negligible thermal mass[3]. Exact surface temperature can be measured only when the surface of interest, the surrounding and the instrumentation are in thermal equilibrium. Any thermal gradient will induce measurement errors as a result of flow of heat. In the Aerospace application, achieving thermal equilibrium is difficult due to design limitation and transient temperature measurement. Hence, the sensor output represents a balance of the following conditions.

        • Heat transferred to the sensing element

        • Heat stored by the element

        -amount of heat lost to the surrounding environment.

        For transient surface temperature measurement the information from the sensor need to be corrected using dynamic calibration data [4].However, location of the temperature sensor, attachment method and test methodologies affect the validity of the data [5].Miniaturized thermocouples are preferred in thermoplastic process to measure temperature in the depth and at the surface of the injected thermoplastic parts due to its less conduction error [6].

        Figure 1. Temperature measurement chain

        Measurement chain contains thermocouple, thermocouple connector, interface cable, signal conditioner and data acquisition system [7]. Cement on thermocouple is preferred in surface temperature measurement due to its flexibility to bond even on curved surface and less perturbation of surface on which temperature to be measured. Proper selection of thermocouple connector and extension wire suitable to the thermocouple is required to minimize the error [8]. Extension wire is a cable used to interface and carry thermocouple output to the signal conditioner. Thermoelectric characteristics of the selected extension wire must be as close as to that of the thermocouple being planned for measurement [9]. Signal conditioner is an interface between thermocouple and Data Acquisition System (DAS). It provides reference junction, compensation, linearization, amplification and band limitation to record correct data. Data Acquisition system is used to monitor/ record thermocouple output.

    2. Construction of Cement-on Thermocouple

      Cement-On thermocouple is a foil sensor. It is embedded between two papers thin; glass reinforced high temperature polymer laminates. Thermocouple leads are electrically insulated from each other as well as being mechanically supported. Lamination also provides a flat surface for cementing. Glass braid insulated 30 gage thermocouple wire is bonded to the foil and strain relieved by laminate to minimize the error due to thermocouple junction interface. It is suitable for bonding on curved and plain surfaces. The polymer/glass laminate limits the maximum temperature of the thermocouple up to 260°C continuous and 370°C for short time about 10 Hours. The response time of the thermocouple varies from 5 to 300 milli seconds depending on the style of construction [10].

      Figure 2. Foil sensor

    3. Principle of Thermocouple

      Surface Temperature Calibrator is preferred to calibrate cement on thermocouple due to its uniformity of temperature on calibrator surface and multisensor calibration feature [12]. Calibration range varies from ambient temperature to 3000 C with an accuracy of ± 10C. It has a built in reference sensor to display the calibration surface temperature. Calibration procedure as per ASTM E220 standard is preferred to maintain traceability [13].

  2. SOURCES OF ERRORS

    Various attributing elements/ factors introduce errors in the surface temperature measurement of the Aerospace components during testing of Aerospace vehicle for its performance evaluation [14].

    Figure 3. Principle of thermocouple

    Thermocouple works on the principle of Seebeck effect. When two dissimilar metals are joined together to form two junctions and if temperature gradient occurs between those two junctions then an emf is generated proportional to the temperature gradient provided the circuit is open [11]. In Fig.3, one of the junction is kept in ice bath ( Cold/Reference Junction) and another junction is fixed to the target surface for measuring temperature (Hot/Measuring Junction).Thermocouple produces output voltage proportional to the difference of temperature between measuring and reference junctions [9].

    1. Calibration of Thermocouple

    Figure 4. Surface temperature calibrator

    • Thermocouple

    • Extension wire

    • Connector Temperature

    • Temperature gradient of T/C connectors

    • Measuring instruments error

    • Bonding/Installation error

    • Polarity reversal

  3. Extension Wire

    Extension wires/compensation wires are used between measuring junction and the reference junction to avoid multi junction formations which act as additional EMF generators. It is necessary to select extension wire and connectors having same thermo electric characteristic as the thermocouple for the entire temperature range. Generally, copper cable and compensating cable are used to interface thermocouple to signal conditioner if the reference junction is a long distance away from the measuring junction. In some cases, thermocouple wire is used as interface cable where the ambient temperature is adverse in nature.

  4. Thermocouple Connector Temperature

    Thermocouple Connector needs to be selected according to the measuring thermocouple to interface thermocouple wire to extension cable without introducing additional EMF genertors. The temperature of the connector may affect the measurement if proper extension wire is not selected.

  5. Temperature Gradient of Connector Terminals

    It is generally assumed that a negligible temperature difference exists across the connector terminals but, in aerospace applications, if proper care in not taken to thermally Isolate thermo connectors from the test article, possibility of temperature gradient between connector terminals occurs and affects the measurement. There are two possibilities on which temperature gradient across connector terminals exist. Connector either body of the connector near to negative terminal touches the body of the test article and positive terminal at ambient temperature or body of the connector near to positive terminal is in contact with the body of the test article and negative terminal at ambient.

  6. Polarity Reversal of Connector Terminals

    When the polarity of the wires is reversed, it causes an error related to the ambient temperature. Although it would be detectable, a double reversal could likewise a cause an error which may not be so easily noticed.

  7. TEST RESULTS AND DISCUSSIONS Surface Temperature Calibrator is used to explore the

possible errors due to compensation cable and copper cable when the interface junction is subjected to be different temperature. Three temperature measurement chains with different extension wires between thermocouple and signal conditioner were configured and calibrated to study the effect of interface cable on temperature measurement due to interface junction temperature. Thermocouple interface connections (thermocouple connectors) were kept in thermal chamber to vary the ambient of thermocouple connectors from

-30°C to 100°C.

TABLE 1. CALIBRATION DATA WITH CONNECTOR AT 25°C

T°C

T1°C

T2°C

T3°C

25

26.2

25.6

4.8

50

51.6

51.4

30.4

100

101.0

101.6

81.6

150

151.5

150.3

133

200

201.8

201.4

180

T- Calibrator Temperature

T1- Configuration with thermocouple wire T2- Configuration with compensating cable T3- Configuration with copper cable

From Table.1, it is clear that the calibration results of three configurations are within the specified error limit when the thermocouple connectors are kept in ambient temperature around 25°C.

TABLE 2. CALIBRATION DATA WITH CONNECTOR AT 50°C

T°C

T1°C

T2°C

T3°C

25

25.4

25.8

85.2

50

50.8

50.5

110.3

100

100.4

99.6

159.2

150

150.2

149.8

210.1

200

201.2

199.5

258.2

T°C

T1°C

T2°C

T3°C

25

25.8

25.4

26

50

51.2

51.6

51.8

100

99.8

101.7

101.4

150

150.4

150.8

150.2

200

201

201.4

202.2

TABLE 3. CALIBRATION DATA WITH CONNECTOR AT 70°C

TABLE 4. CALIBRATION DATA WITH CONNECTOR AT 100°C

T°C

T1°C

T2°C

T3°C

25

25.8

26.1

-16.6

50

51.3

51.7

+8.2

100

101.6

101.2

56.8

150

150.8

150.6

105.8

200

201.4

201.7

158.8

From the tables 2-4, copper cable is not suitable for temperature measurement if the connector temperature exceeds 40°C. In case of T3 configuration, measurement error increases as the thermocouple connector temperature increases. But the calibration data pertaining to other two configurations (T1&T2) are within the tolerance limit irrespective of thermocouple connector temperature.

TABLE 5. CALIBRATION DATA WITH CONNECTOR AT-30°C

T°C

T1°C

T2°C

T3°C

25

25.6

25.4

-34.8

50

50.6

49.4

-6.2

100

98.8

99.4

+42.4

150

150.7

151.2

88.7

200

201.2

199.7

141

From table 5, it is evident that T3 configuration read more than the calibrator temperature when thermocouple connector temperature is -30°C.But other two configurations read close to the calibrator temperature.

Figure 5. Experimental set up

An experiment was carried out on three temperature measurement chains of configurations as shown in Fig.5 to analyze the error due to the temperature gradient across thermocouple connector terminals. Fig.6-8 displays the deviation of temperature data among the three configurations.

Temperature in deg.C

150

100

+ side of connector touching body connector not touching the body

50

0

0 50 100 150 200 250 300

Time in seconds

Figure 6. Deviation due to connector side (+) touches body

Temperature in deg.C

Connector not touching body

– side of connecotor touching body

140

120

100

80

60

40

20

0 50 100 150 200 250 300

Time in seconds

Figure 7. Deviation due to connector side (-) touches body

Calibration of thermocouple with reverse polarity was carried out using surface temperature calibrator to determine the formula for correcting the error due to polarity reversal. From the calibration results shown in table 6, correct data can be achieved from polarity reversal thermocouple data using the Eqn. (1)

Tc = 2* Ta – Tp (1)

Tc – Corrected data (°C)

Ta – Ambient temperature (°C)

Tp – Polarity reversed thermocouple data (°C)

TABLE 6. POLARITY REVERSED THERMOCOUPLE DATA

Calibrator Temperature

°C

T1°C

(without Polarity Reversal)

T2°C

(with Polarity Reversal)

T2°C

(corrected Data)

30

30

30

30

60

60

0

60

90

90

– 30

90

120

120

– 60

120

As a case study, an additional temperature channel (thermocouple) with polarity reversal of wire was recorded along with other temperature channels during static testing of aerospace vehicle. Fig. 9 shows the deviation between the polarity reversed thermocouple test data and near location test data.

150

Polarity reversal data Near location data

300

Temperature in deg.C

100

50

0

+ side of connector touching body connector not touching the body

– side of connector touching the body

200

Temparature in Deg.C

100

0

-100

-200

0 20 40 60 80 100

0 50 100 150 200 250 300

Time in seconds

Figure 8. Deviation due to connector side (+/-) touches body

Time in Seconds

Figure 9. Polarity reversed thermocouple data profile

300

Temperature in deg.C

200

100

0

-100

-200

Corrected data Polarity Reversal data

may not affect the measurement accuracy as long as temperature gradient across connector terminals is zero. As any temperature gradient leads to inaccuracy of measurement, it is necessary to thermally isolate junction of thermocouple connector and extension wire from adverse temperature in testing of aerospace engines. Though, correct data can be derived from polarity reversal data, it is recommended to avoid polarity reversal by heating/cooling the thermocouple junction before test. However, double polarity reversal cannot be identified easily; care must be taken while interfacing thermocouple to signal conditioner.

REFERENCES

Figure 10. Corrected data profile

Fig.10 shows the profile of corrected data using Eqn.1. Corrected data profile is a mirror image of polarity reversed data about the ambient temperature. Fig.11 highlights the closeness of corrected data with respect to near location data.

300

Temperature in deg.C

250

200

electronic packages,IEEE Semi- Therm 19th symposium, pp.97- 105,2003 [0-7803-7793-1-02@2003 IEEE].

0

20

40 60

80

100

[18]

H.Shaukatullah and Alan Claassen, Effect of thermocouple wire size

Time in seconds

and attachment method on measurement of thermal characteristics of

  1. Hemanshu D Bhatt,Ramakrishna Vedula,Seshu B. Deshu and Gustave

    C. Fralick, Thin film TiC/TaC thermocouples,Thin solid films, 342 (1999) 214-220.

  2. Otto J. Gregory,Eike Busch,Gutsave C Fralick and Ximing Chen Preparation and characterisation of thin film thermocouples , Thin solid films, 518 (2010) 6093-6098.

  3. Zhou Hanchang,Hao Xiaojian,Chen Weili and Huang Liang,Traceable dynamic calibration for high temperature sensors using co2 laser, IEEE Sensors, pp.1349-1352,EXCO, Daegu,Korea,October 2006 [ 1-4244- 0376-6-06@2006 IEEE].

  4. Robert L. Kozarek,Effect of case temperature measurement errors on

    the junction-to- case thermal resistance of a ceramic PGA, IEEE Semi- Therm seventh symposium,pp.44-51,1991[CH2972-8-91-0000-

    150

    100

    50

    0

    Corrected data

    Near location data

    0044@1991 IEEE].

  5. Youssef Farouq, Cecile Nicolazo, Alain Sarda and Remi Deterre, Temperature measurements in the depth and at the surface of injected thermoplastic parts , Measurement 38 (2005) 1-14.

  6. K. Fowler and J. Schmalzel, Sensors: The first stage in the measurement chain, IEEE Instrumentation & Measurement magazine,

    0 20 40 60 80 100

    Time in seconds

    Figure 11. Comparison of corrected and near location data profile

    CONCLUSION

    Temperature measurement errors due to interface elements with different possible configurations and environments have been analyzed with experimental results. Extension wire and thermocouple connector suitable to thermocouple need to be selected to minimize the error due to additional EMF generators. Copper cable can be used as an extension wire provided the thermocouple connector temperature is below 30°C .Incase, compensating cable/thermocouple wire is used as extension wire, connector temperature up to 100°C,

    vol 7, pp. 6066, Sept. 2004.

  7. Nanmac corporation temperature measurement hand book,Vol.

    X11,pp.L3-L6 , 2003

  8. Manual on the use of thermocouples in temperature maesurement , American Society for Testing and Materials, STP470B,1981 .

  9. Practices of temperature measurement, Omega Engineering ,2006.

  10. Thomas J. Bajzek, Thermocouples: A sensor for measuring temperature

    , IEEE Instrumentation & Measurement Magazine, pp. 35-40, March 2005 [1094-6969-05@2005 IEEE].

  11. Hart scientiic temperature calibration equipment catalogue,pp.135- 136,2003-2004.

  12. ASTM- E220 Stanadard

  13. L de Vries-Venter and GP Hancke, Validating surface temperaure measurement, IEEE Instrumentation and Measurement Technology conference,pp.1381-1385,1997 [0-7803-3312-8-97@1997 IEEE]