Method and Model for Jaundice Prediction Through Non-Invasive Bilirubin Detection Technique
Ankan Gupta(1)
Department of Electronics and Communication Engineering Ambala Collage of Engineering and Applied Research Ambala, India
Ashok Kumar(2), Preeti Khera(2)
Department of Electronics and Communication Engineering Ambala Collage of Engineering and Applied Research Ambala, India
Abstract In this paper, we have focused on developing a method and model to detect jaundice using non-invasive techniques. The yellow discoloration is used to measure bilirubin concentration for determining the level of jaundice in infants. Hyperbilirubinemia or Neonates Jaundice is commonly happened in the neonates or newborns due to rise in the of the amount of bilirubin concentration in the body. Current detection techniques, however, require clinical tests with blood samples or other particular equipments. As a result, newborns often depends on visual assessments of their skin color at home, which is known to be unpredictable. Therefore, to detect newborn jaundice non- invasively we are going to develop a technique based on yellow discoloration of the skin. Neonatal hyperbilirubinemia, especially kernicterus; may be prevented by screening for neonatal jaundice at early stage. The proposed system is accurate in the diagnosis of neonatal jaundice, and can reduce the need for blood sampling.
KeywordsHyperbilirubinemia; non-invasive; Transcutaneous Bilirubinometer; Transcutaneous Serum Bilirubin
-
INTRODUCTION
Numbers of medical devices are becoming increasingly common for fitness, heart rate check and various other devices. As the technology increasing day by day, people move towards the efficient and low cost technology. People want the technology that save time money and provide good results.
Neonatal jaundice is the common problem in the newborns [1,2]. It is found that the neonatal jaundice occur more in Asian neonates than the Caucasian neonates. Jaundice is also known as icterus. It causes yellowing of skin and whitening of eye that is known as sclera. Almost all the [3] infants are jaundice proned; among them around 60% are term and 80 % are preterm infants. Jaundice occurs due to breakdown of Red Blood Cells; the breakdown process is known as Hemolysis if the cell breakdown rate occurs at rate faster than the usual it increase the level of bilirubin in the body and causes jaundice to the infants. Bilirubin concentration can be detected using two technique i.e. Invasive techniques and Non-invasive techniques. In invasive detection, a physical examination is carried out to look for sign of swelling of liver and legs ankles or feet, which might indicate cirrhosis of the liver. Urine be tested to check Urobilinogen, which is produced when [3] bilirubin is broken down to find high or low level of bilirubin concentration can help pin point the type of jaundice. Blood test may be used. Whereas in Non-invasive detection can be
done by using Trancutaneous bilirubinometer. Transcutaneous Bilirubinometry works by directing light into the skin of neonate and measures the intensity of specific wavelength that is returned. The number of wavelengths, used is variable in different transcutaneous bilirubinometer. The meter analyzes the spectrum of optical signal reflected from the neonates subcutaneous tissues. These optical signals are converted to electrical signal by a [4] photocell. These are analyzed by a microprocessor to generate a serum bilirubin value.
A moderate level (6mg/dl- 9 mg/dl) can be tolerable and easily treated but delay in the treatment can cause fatal and irreversible brain damage of the newborn. Therefore, early detection of the newborn jaundice is mandatory. Most of the infants are affected by jaundice after discharged from the hospitals. Visual assessment generally used at home, where laboratory takes too much time to provide the results. Parents and clinicians are usually determine the jaundice by visual assessment. Various studies show that the accurate results are not provided by visual assessments even experienced professionals are unable to determine the exact level of jaundice. In this paper we present the design of a system that evaluate jaundice level by detecting serum bilirubin coloration on a strip. The system is based upon the smartphone embedded camera images and a calibration card formed in the Matlab.
-
RELATED WORK
James W. Kronberg et al. [4] introduced a trancutaneous bilirubin detection device. This device was based on the absorption of light by the skin at particular wavelength. The amount the light absorbed by the skin is proportional to the amount of illness. Steven L. Jacques et al. [9] analyzed a
;device, which was based upon the maturity dependent optical properties of the skin. Refraction of red , yellow orange lights along with these light reflection of blue light determine the concentration of bilirubin. Later on Buttitta et al. [10] introduced a device in which two lights were transmitted through the infant skin one is completely absorbed and another one was partially or not absorbed by the skin. The reflected light intensity is used to determine the bilirubin concentration. David P. Dewitt et al. [8] developed a device which was also based on the two-filter design technique. This device was able
to determine the type of the treatment required to prevent the hyperbilirubinemia. Vinod K Bhutani et al. [11] introduced a device that depends upon the multi-wavelength spectral reflectance analysis. This device concentrates on those predischarge infants who have high risk of hyperbilirubinemia. Giovanna Bertini et al. [12] Introduced a first electronic device, which was used to determine whether neonates required TSB detection, or not. Gagan Mahajan et al. [13]study was undertaken to relate the correlation between the transcutaneous bilirubinometer with total serum bilirubin reading without using phototherapy. Yu-Hsun Chang et al. [2] evaluated the usefulness of JM-103 in Taiwanese neonates. At value of 9.4 mg/dl this device showed significant detection. Brad S. Karon et al. [14] determined that transcutaneous bilirubinometer able to detect risk of hyperbilirubinemia. Samar N. El-beshbishi et al. [1] discussed the review that the Transcutaneous bilirubinometer could be used for early detection of the sever hyperbilirubinemia. M. Penhaker et al.
[15] proposed a device for detection of bilirubin. This device is ten times cheaper and easy to use. The variation of error was less than 4%. KA Jangaard et al. [16] Proposed that heel punctured cause painful to infants and distress to the parents. Billicheck can be used for detection of neonatal those who didnt receive phototherapy and does not have any patch or mark on the skin. Nurashlida ali et al. [17] introduced a non- invasively detection method for jaundice. This method uses optical techniques for detection. This technique is better to eliminate the pain occurred to the infants during taking blood samples.-
Gaps
-
Transcutaneous bilirubin instrument measurements were inaccurate for measuring bilirubin levels in infants those were already receiving phototherapy and if an area of skin has any mark or patch.
-
The mechanism was not as sensitive in the minute sample of preterm infants, and a larger reading was required before recommending the use of this instrument in this inhabitants.
-
Transcutaneous bilirubinometer also affected by the color of the skin and the hemoglobin concentration present in the body because the color of the skin and hemoglobin concentration varied from person to perso so that the measuring values are also varied.
-
The biggest gap is that tracutaneous bilirubinometer is not able to detect the Hyperbilirubinemia (i.e.>= 14 mg/dl) stage correctly, which is the dangerous stage of jaundice.
-
-
-
ANALYSIS
To determine whether a newborn should receive phototherapy or an exchange blood transfusion, doctors or nurses reference specialized graphs with the newborns age, number of weeks of gestation, and bilirubin [11] level. Figure
(1) explain Bhutani nomogram graph. The Bhutani nomogram provides a means to evaluate a newborns risk based on the percent of newborns in the study with given bilirubin levels and ages. High intermediate risk is considered above the 75th percentile, and high risk above the 95th percentile. Bilirubin
levels are commonly expressed in milligrams per deciliter (mg/dl) or micromoles of bilirubin per liter (mol/L)
Bhutani Nomogram
20
15
10
5
0
Low Risk
Medium Risk High Risk
12 36 60 84 108132
Postnatal Age (hours)
Serum Bilirubin (mg/dl)
Fig. 1. Bhutani Nomogram Graph for Standard Bilirubin Concentration
-
Experimental Method to Measure Bilirubin Concentration Non-Invasively
Feature Extraction
Color Balancing
Take an Image
The bilirubin concentration can be measured non- invasively by determining the yellow color discoloration of the skin. The steps to calculate the bilirubin concentration is described as below in the figure (2):
Bilirubin Estimation
Apply Regression
Fig. 1. Steps to Calculate the Bilirubin Concentration
-
Color Balancing
The image was converted into HSV model and normalized value of Red, Green, Blue was computed to make some of the effects of different lightning condition less [20] severe. The HSV stands for the Hue, Saturation and Value.
As hue varies from 0 to 1.0, the corresponding colors vary from red, through yellow, green, cyan, blue, and magenta
, back to red, so that there are actually red values both at 0 and
1.0. As saturation varies from 0 to 1.0, the corresponding [20] colors (hues) vary from unsaturated (shades of gray) to fully saturated (no white component). As value or brightness,varies from 0 to 1.0, the corresponding colors become increasingly brighter. Compute R,G,B and cmax, cmin value for RGB to HSV conversion of the image as done using equation (1) to (13):
R= R/255 (1)
G= G/255 (2)
B= B/255 (3)
cmax= max(R,G,B) (4)
cmin= min(R,G,B) (5)
= (6)
Hue Calculation:
= 00 , = 0 (7)
= 600 ( 6) , = (8)
= 600 ( + 2) , = (9)
= 600 ( + 4) , = (10)
Saturation Calculation:
= 0, = 0 (11)
-
Least Angle Regression- LARS regression uses a variant of forward feature selection to decide what features are most useful. Instinctively, this regression helps to eliminate redundant features, while [2] creating new features based on their association to the chosen features. Essentially, the best predictor from the feature set is chosen by developing a single-feature, linear regression
[15] from each feature. -
Lasso – Elastic Net Regression- Lasso is a regularization technique for performing linear regression. Lasso includes penalty term that constrains the size of the estimated coefficients. Lasso is a shrinkage estimator; it generates coefficient estimates that are biased to be small. A lasso estimator [2] can have smaller mean squared error than an ordinary least-square estimator when it is applied it to new data. Lasso sets more coefficients to zero. Elastic net is a hybrid of ridge regression and lasso regularization. Like lasso, elastic net can generate reduced model by [13] generating zero-valued coefficients as described by equation (13). For given value of a non-negative
=
, 0 (12)
parameter, lasso solves the problem of non- negative parameters.
Value Calculation:
min 1 N
(yi
xT) + p
| |
(13)
o 2N
j=1 0 i
j=1 j
= (13)
-
Color Chart Formation
The color chart was created for standard bilirubin level
[15] shades of each block size of 26 *26 and one block was kept blank. Where, x represents the empty/ hollow block to paste strip/skin sample image as shown in figure (3)x
Fig. 3 Color Chart
-
Machine Regression
The regression algorithm was applied on the card, which contains the sample image pasted on one block. The regression algorithm employs an ensemble of different regression. The aim of [2] machine learning is to build a model that makes decisions based on facts in the presence of uncertainty. As adaptive algorithms identify [13] patterns in color segment. A computer "learns" from the observations, therefore, when exposed to more clarification, the computer improves its decision-making performance. In this model four different Regressions are used. The four regression are
-
-
K-NN Regression- The first regression algorithm is an encapsulated k-Nearest Neighbor regression (k = 7) [17]. Intuitively, this regression takes a more local estimate of the bilirubin level based upon [9] training points that have similar feature values. In this regression, when an unknown test vector is analyzed, the k-nearest neighbors are found around the test vector in the [2] database of features. The L1 norm is a variable, which is used to calculate the nearest neighbors feature. Feature points from the neighbors are used to train a linear support vector regression and new regression is built each time whenever a new test point is analyzed.
-
Support Vector Regression- Instead of minimizing the observed training error, Support Vector Regression attempts to minimize the generalization error bound so as to accomplish generalized performance. The idea of SVR is based on the computation of a linear regression [31] function in a high dimensional feature space where the input data are mapped via a nonlinear function. Support Vector Machine are learning machines implementing the structural risk minimization inductive principle to obtain good generalization on a limited number of learning patterns. Structural risk minimization involves concurrent attempt to minimize the empirical risk.
-
Final Output
If the difference is less than the empirically derived threshold of 1.0 mg/dl, the collection agrees and the mean is chosen. If the difference is greater than 1.0 mg/dl, then the second highest bilirubin value (i.e., the 90th percentile) is chosen. This helps to bias the regression algorithm to select a large bilirubin value when the ensemble does not agree. If the variation of experimental value of bilirubin concentration from the actual value of bilirubin concentration is greater than 5% then the technique is not considered to be completely accurate.
Therefore, the developed technique variation must be less than the 5 % of the actual value then to validate the result. Table 1 contains the variation of actual and experimental result and the error percentage due to variation in output from standard value of bilirubin concentration.
Observ ations |
Bilirubin Concentration |
||
Actual Values |
Experimental Values |
Error Percentage |
|
1 |
3 |
3 |
0 |
2 |
6 |
6.2 |
3.3 |
3 |
9 |
9 |
0 |
4 |
12 |
12 |
0 |
5 |
15 |
<>15 |
0 |
6 |
18 |
19.5 |
8.3 |
7 |
21 |
22.7 |
8.09 |
8 |
24 |
26 |
8.33 |
9 |
27 |
29.5 |
9.2 |
TABLE I. TABLE STYLES
The experimental results were able to detect jaundice for bilirubin concentration upto 14mg/dl where as our work able to detect jaundice for bilirubin concentration level of 17mg/dl.
FUTURE SCOPE
Jaundice detection at early stage is necessary to avoid high risk of hyperbilirubinemia. So that it is required to detect jaundice at early stage. The future work can be extended to detect jaundice for values greater than 17mg/dl and can be implemented for different skin shades. A app can be develop for smartphones so that jaundice can be detect at home at early stage without need to rush towards the laboratory to detect jaundice
Variation
Figure (3) describes the variation of experimental value from actual value of bilirubin concentration and its error percentage
35
30
25
20
15
10
5
0
3 6 9 12 15 18 21 24 27
Bilirubin Concentration
Actual Value
(av)
Experimental Value (ev)
Error %
Fig.3. Error Percentage CONCLUSION
We have designed a new method and model to detect jaundice, which is based on the yellow discoloration of the bilirubin samples. The outputs of the different bilirubin samples are included in the result. The R,G,B value is varied for different concentration of the bilirubin. We take different samples of bilirubin in different condition of light. Firstly the 3mg/dl sample has been taken and then compared it with the standard color chart, it is analyzed that the exact value of the sample is determined by the developed technique. We repeat this process for 6mg/dl, 9mg/dl, 12mg/dl, 15mg/dl, 18mg/dl, 21mg/dl, 24mg/dl, and 27mg/dl bilirubin concentration sample image, we find that the sampled value up to 17 mg/dl the developed device is able to detect bilirubin concentration reliably. After 17mg/dl, the developed technique is not able to detect bilirubin concentration exactly. The error percentage is less than 5 % for bilirubin concentration value up to 17mg/dl.
REFERENCES
-
Samar N.El-Beshbishi, Karen E. Sharttuck, Amin A. Mohammad and John R.Petersen,Hyperbilirubinemia and Transcutaneous Bilirubin- ometry,clinical chemistry, vol. 55, no.7, 2009, pp. 1280-1287.
-
Yu-Hsun Chang, Wu-Shiun Hsieh, Hung-Chieh Chou, Chien-Yi Chen, Jing-Yi Wu, Po-Nien Tsao The Effectiveness of a Noninvasive Transcutaneous Bilirubin Meter in Reducing the Need for Blood Sampling in Taiwanese Neonates,clinical Neonatalogy.,vol.13, no. 2, 2006, pp. 60- 63
-
Smriti Shrivastava,Diagnosis of Neonatal Jaundice Using Artificial Neural Networks,International Indexed & Refereed Research Journals, vol. 4, 2013, pp. 69-72.
-
James W. Kronberg. Optical Transcutaneous Bilirubin, US Patent 5 259 382, November 9, 1993.
-
Tai-Wing Wu, Rochester, N.Y., Assay For Bilirubin, United State Patent 4 069 016, January 17, 1978 pp.1-32.
-
Moe Akahira-Azuma, Naohiro Yonemoto, Battsengel Ganzorig, Rintro Mori, Shinichi Hosokawa, Takeji Matsushita, Bayasgalantai bavuusuren, validation of Trascutaneous bilirubin meter in Mongolian Neonates: comparison with total serum bilirubin,BMC Pediatrics. Vol. 13, no. 151, 2013, pp. 1-7.
-
Mark McEwen, Karen Reynolds, Noninvasive Detection of Bilirubin Using Pulsatile Absorption,
-
David P. Dewitt, Robert E.Hannemann, John F. wiechel Method For Determining Bilirubin Concentration From Skin Reflectance. United States Patent. 4 029 085, june 4, 1977 pp. 1-8.
-
Steven L. Jacques, David G. Oelberg, Iyad Saidi, Method And Apparatus For Optical Measurement Of Bilirubin In Tissue. United States Patent. 5,353,790, 1994, pp. 1-79.
-
Buttitta, Anthony D, Librtyville, Illinois, Falk, Steven M.Silver Spring, Maryland, Non-Invasive Bilirubin Monitor. European Patent Application. Ep 0 747 002 A1 , December 11,1996, pp.1-11.
-
Vinod K. Bhutani, Glenn R. Gourley, Saul Adler, Bill Kreamer, Chris Dalin and Lois H. Johnson ,Noninvasive Measurement of Total Serum Bilirubin in a Multiracial Predischarge Newborn Population to Assess the Risk of Severe Hyperbilirubinemia,the American Academy of Pediatrics. Vol. 106, 2000, pp. 1-9.
-
Giovanna Bertini,Non-invasive bilirubinometry in neonatal jaundice.
Semin Neonatol. volume 7, 2002, pp.129-133
-
Gagan Mahajan Transcutaneous Bilirubinometer in Assessment of Neonatal Jaundice in Northern India. INDIAN PEDIATRICS. Vol. 42, 2005, pp. 41- 45
-
Brad S. Karon, MD, Ann Teske, Paula J. Santrach, MD and Walter J. Cook, Evaluation Of The Bilichek Noninvasive Bilirubin Analyzer For Prediction Of Serum Bilirubin And Risk Of Hyperbilirubinemia. American Journal Clinical Pathology. Vol. 130, 2008, pp. 976-982.
-
M. Penhaker, V. Kasik, B. Hrvolova Advanced Bilirubin Measurement by a Photometric Method Elektronika Ir Elektrotechnika. Vol.19, 2014
, pp. 47-50.
-
KA Jangaard. Estimation of bilirubin using BiliChek, a Transcutaneous bilirubin measurement device: Effects of gestational age and use of phototherapy. Paediatr Child Health. Vol. 11 no. 2, 2014, pp. 79-83.
-
Nurashlida Ali A Review of Non Invasive Jaundice detection using
Optical Technique in Neonates.Conference in Advances In Computing, Electronics and Electrical Technology – CEET 2014. , 2014,pp.1-3.
-
Mark J.Rice, Robert H.Sweat, Jr. James M. Rioux, Williams T.Williams, Wilson Routt, Non-Invasive Measurements of Blood Component Using Retinal Image, United State Patent US 6 305 804 B1, October 23, 2001 pp. 1-10
-
Hetal Vora, Pankti Desai, Invasive Versus Non-Invasive Hemoglobin Measurements by Pulse Co-Oximeter in Neonates Admitted To NICU, NHL Journal of Medical Science, vol. 3, January 2014, pp.68-71.
-
Durgesh Nirapure, Udaypal Reddy,Fast Retrieval of Images using Filtered HSV Color Level Detection, International Journal of Emerging Technology and Advanced Engineering, Volume 3, Issue 8, 2013, pp. 414-419